2D Laser Blood Flow Imager
OZ-EX1 – Two-Dimensional Laser Blood Flow Imaging System
OMEGAZONE OZ-EX1 · Laser Speckle Imaging · For rats, mice and small animal research
1. Overview
The OZ-EX1 is a two-dimensional laser blood flow imaging device designed for measuring tissue blood flow in small animals such as rats and mice. It enables non-invasive, nearly continuous, high-speed, and high-resolution measurement of blood flow distribution in tissues—including the brain and lower extremities—and displays the magnitude of blood flow as a color gradient.
The measurement software includes the following three modes:
Measurement Modes
Mode | Frame Rate | Image Resolution |
1. High-Resolution Mode (Hi-Res) | Approximately 1 frame per second | 639 × 480 |
2. Hi-Speed Mode | Approximately 15 images/second | 213 × 160 |
3. Fast Average Mode (HS-AVG) | — | 213 × 160
|
Since blood flow at each point is measured simultaneously, it is possible to determine both the blood flow correlation at each point and the variations in blood flow. The measurement range can be adjusted over a wide range – from approximately 8 mm × 6 mm to approximately 108 mm × 80 mm – simply by changing the lens. The software is divided into an LSI (for measurement) and a LIA (for analysis), enabling multiple analyses of the measured and stored data after completion of the measurements.
The OMEGAZONE OZ-EX1 is a compactified version of our OZ-2 two-dimensional laser blood flow imaging system, which integrates the camera and laser unit into a more compact design. This software is a revised version of the OZ-2 software; therefore, its core functionality and operating procedures remain identical to those of the OZ-2.
2. Working Principle
When biological tissue is irradiated with a laser and a photodetector is used to detect scattered light from stationary tissues and scattered light from flowing red blood cells. When light scatters from a cell, the light intensity varies due to interference change. This two-dimensional interference pattern is called a speckle pattern, and the speckle intensity degree varies over time in accordance with the number and flow rate of red blood cells. As the flow velocity increases, the rate of change in light intensity also increases; even with an increase in the number of photons, this rate of intensity change accelerates.
Doppler frequency shift effect caused by red blood cell movement |
Measuring Principle
3. Features
- The laser unit and camera/lens combination are compact, enabling measurements to be performed in confined spaces.
- The power supply for the laser unit and the camera is provided by the computer's USB port; therefore, the external power source is solely for powering the computer.
- The software is divided into two components: LSI for measurement and LIA for analysis.
- Color mode: Grayscale / Blue-Red.
- Marking Function: The feature to add markers to images and charts during an event.
- Differential display: The increase or decrease in blood flow is determined by comparing the displayed image with the reference blood flow image.
- Zero-cutting operation: Calculate the average blood flow value for the ROI-enclosed region, excluding zero values.
- ROI Saving Feature: You can save and retrieve up to 8 ROIs.
- Blood flow value: The average blood flow value for the region surrounding the ROI can be saved as a CSV file.
- Automatic Gain / Manual Gain: Image brightness can be adjusted automatically or manually.
- Optical Zero: A feature that reduces image brightness to zero when no laser beam is being applied, enabling measurements to be performed without the influence of external light.
- Two color modes: Standard/Enhanced.
- Marking Function: The feature to add annotations to images and charts during an event.
4. Specification
Instrument
Item | Detail |
Model | OZ-EX1 |
Operational principle | Laser speckle imaging |
Equipment Composition | Instrument, laser device, USB interface CMOS camera, stand, computer |
Hardware — Laptop-based image processor (instrument)
Item | Detail |
Operating system | Windows 11 |
Size | 380 x 40 x 250 (W x H x D) |
CPU | Intel iCore, or AMD Ryzen |
Weight | 3 kg |
Internal storage | 8 GB or higher |
DVD | Multi-driver system (supports external devices) |
Screen size | 14 x 16 inches, 1920 x 1080 |
Source | 220 V / 50 Hz |
Software
Item | Detail |
Software image pixel count | 1600 x 900 |
Power | 350 w |
Weight | 35 kg |
Size | 350 x 200 x 520 mm (W x H x D) — Support |
Support
Item | Detail |
Size | Height: 30 cm × Arm length: 15 cm |
Light source power supply | — |
Weight | 1.5 kg |
Source | USB 5 V, 0.4 A |
CMOS Camera
Item | Detail |
Size | 16 x 6 x 15 (W x H x D) |
Imaging time | 1/30 sec |
Weight | Approx. 0.5 kg |
Signal output mode | USB, 8 bit |
Measurement Laser
Item | Detail |
Wavelength | 780 nm (laser) |
Power | ≤ 70 mW; CLASS 1 |
5. Software Functions (LSI measure / LIA analyse)
Function | Description |
FLOWprice | Average value within the ROI region. The change in blood flow values over time is displayed in the bottom-left panel. |
ROI | The function for defining a survey area: LSI allows the definition of two areas, while LIA allows the definition of up to eight areas. |
MAX | Set the upper limit of the measurement range for coloration. |
MIN | Set the lower limit for the coloration measurement range. Areas where the measurement result is below this value are displayed in gray or blue. |
Pattern | HR, HS, HSA |
Data | The measured blood flow values within the selected ROI can be exported and saved as an Excel file. |
Mode 1 — High Resolution | High Resolution (High-Resolution Mode): 1 frame / second, Resolution: 639x480 |
Mode 2 — Hi-Speed | Hi-Speed (High Speed Mode): 15 frames/second, Resolution: 213x160 |
Mode 3 — HS-AVG | HS-AVG (Fast Average Mode): Set the average, resolution: 213x160 |
Histogram | The histogram of blood flow is displayed in the bottom-right corner. |
ZE Mode – Zero Abnormality Mode | A function that calculates blood flow by excluding the zero-value portions from the selected region. |
Play animation | Provides a feature to play saved consecutive images, allowing you to set the number of images to play per second. |
Save Settings | Setpoint Preservation Function |
Forward & Backward | Display the previous or next image |
Initial Settings | Return each function to its original function setting. |
Previous & Next Jump | The "Jump to previous / next few images" feature allows you to specify the number of images to skip. |
6. Applications
- Cortical blood flow measurement
- Measurement of spinal cord and peripheral blood flow
- MCAO molding
- Measurement of blood flow during vasodilation, vasoconstriction, and ischemia
- Tissue blood flow measurement
- Skin blood flow measurement
- Gingival blood flow measurement
- Gingival blood flow measurement
- Mucosal blood flow measurement
- Allergy Research
- Environmental Physiology Research
7. Company Information & Contact
Guining (Shanghai) Experimental Equipment Co., Ltd.
©2024 Labcan Scientific – All rights reserved. Labcan is committed to continuously improving its product functionality and reserves the right to modify its products and technical specifications without prior notice; for further details, please consult your local technical engineer.
Labcan Scientific Supplies Shanghai Co., Ltd
Contact
Item | Detail |
WeChat | Feel free to add our WeChat account to learn more. |
Technical Support Hotline | 400 920 2386 |
Email | info@labcan.cn |
Website | www.aschchina.cn |
©2024 Labcan Scientific – All rights reserved.