The OZ-EX1 is a two-dimensional laser blood flow imaging device designed for measuring tissue blood
flow in small animals such as rats and mice. It enables non-invasive, nearly continuous, high-speed, and
high-resolution measurement of blood flow distribution in tissues—including the brain and lower
extremities—and displays the magnitude of blood flow as a color gradient.
The measurement software includes the following three modes:
1. High-Resolution Mode (Hi-Res): Approximately 1 frame per second; image resolution: 639 × 480
2. Hi-Speed Mode: approximately 15 images/second, with an image resolution of 213 × 160
3. Fast Average Mode (HS-AVG): Image resolution – 213 × 160
Since blood flow at each point is measured simultaneously, it is possible to determine both the blood flow correlation at
each point and the variations in blood flow. The measurement range can be adjusted over a wide range – from
approximately 8 mm × 6 mm to approximately 108 mm × 80 mm – simply by changing the lens. The software is divided into
an LSI (for measurement) and a LIA (for analysis), enabling multiple analyses of the measured and stored data after
completion of the measurements. The OMEGAZONE OZ-EX1 is a compactified version of our OZ-2 two-dimensional laser
blood flow imaging system, which integrates the camera and laser unit into a more compact design. This software is a revised
version of the OZ-2 software; therefore, its core functionality and operating procedures remain identical to those of the OZ-2.
When a laser irradiates biological tissue and a photodetector receives scattered light from stationary tissue as well as scattered light from flowing red blood cells, the light intensity varies due to interference.
This two-dimensional interference pattern is called speckle. The speckle intensity changes over time depending on the quantity and flow velocity of red blood cells.
As the flow velocity increases, the fluctuations in light intensity become faster. Such intensity fluctuations accelerate even when the number of red blood cells rises.
Each pixel of the CMOS camera captures these intensity variations and performs computations to render the tissue blood flow values as a two-dimensional color image.
The laser light received by the photodetector is a composite of surface-reflected light reflected off the surface of biological tissue and scattered light containing blood flow signals originating from deep within the biological tissue. The surface-reflected light carries no signal component, yet its reflection intensity itself fluctuates, causing large oscillations in the received light intensity and resulting in artifacts. Since the polarization direction of surface-reflected light matches that of the laser beam, stable blood flow measurement can be achieved by only collecting light polarized perpendicularly to the laser beam.
The camera of the OZ-EX1 is equipped with a polarizing filter to receive only orthogonally polarized light, together with a bandpass filter for the laser wavelength range.
Characteristic
・ The laser unit and camera/lens combination are compact, enabling measurements to be performed in confined spaces.
・ The power supply for the laser unit and the camera is provided by the computer's USB port; therefore, the external
power source is solely for powering the computer.
・ The software is divided into two components: LSI for measurement and LIA for analysis.