Evaluation of Moisturizing and Oil-Controlting Effects of Facial Cleansing Products Based on T/GDCA 008-2021: Application of the HydraSebum Duo Device

Created on 10.20
When conducting human subject evaluations for cleansing, oil-control, hydrating, or barrier-functioning skincare products, one should not rely solely on whether the skin feels "refreshing after cleansing." It is essential to objectively record the following parameters: how quickly sebum levels rebound after cleansing, whether skin hydration levels change, and whether the skin's oil-and-water balance is altered after continuous use. Application amount, massage duration, and cleansing technique are merely testing conditions; the ultimate evaluation must be based on skin biomarker data. The standard T/GDCA 008-2021, "Human Subject Testing Methods for Moisturizing and Oil-Control Performance of Cleansing Products," provides specific human testing protocols for these properties; the EEMCO Guidelines for Electrical Measurement of Skin Hydration further emphasize considerations regarding environmental conditions, test site selection, and operational protocols. The Labodorf HydraSebum Duo skin hydration and sebum analyzer complies with this standard, enabling simultaneous measurement of skin hydration and sebum content, thereby facilitating standardized comparative analyses at fixed measurement sites.
Challenges in facial oil and moisture analysis
1. When relying solely on visual inspection and tactile feedback, it is difficult to make consistent comparisons between the before and after states.
Perceptions of freshness, tightness, oiliness, and softness can be influenced by lighting conditions, environmental factors, and individual subjective experiences; consequently, the same product may be described differently by different users. When conducting efficacy evaluations, recording data such as water content and oil content facilitates comparative analysis over time.
2. Focusing solely on the decrease in oil content can easily lead to overlooking changes in moisture content.
When cleansing efficacy is high, sebum levels may decrease, and skin moisture may also decline, potentially leading to dryness or a tight sensation. Therefore, when evaluating oil control, it is advisable to simultaneously measure skin moisture levels to avoid relying solely on changes in sebum levels.
3. The timing, anatomical site, and pressure levels are inconsistent, making data comparison difficult.
Sebum and water content vary depending on the time of day, ambient temperature and humidity, frequency of cleansing, measurement site, and applied pressure. When conducting a before-and-after comparison, the acclimatization period, measurement site, and applied pressure must remain constant; otherwise, the results will be difficult to compare.

How can the oil and moisture content of facial skin be measured?

1. Subject and Environmental Preparation

Prior to the test, record the skin type, the time of your most recent cleansing session, whether you have used skincare or cosmetic products, and the measurement site. Before the formal test, you may acclimatize for 15–30 minutes in an environment maintained at 20–22 °C with a relative humidity of 40%–60%; measurement should be avoided immediately after exercising, sweating, or cleansing your face.
0

2. Instrument Preparation

Ensure that the device has sufficient battery power, connect it to your computer via Bluetooth, and select the appropriate COM port in the software. Before testing, clean the sensor with a dry, soft cloth to avoid residual water, sweat, skincare products, or oil stains.

3. Test Site and Procedure

Measure skin oil content in a 3 cm × 3 cm area on the forehead, and measure skin water content and transepidermal water loss in a 3 cm × 3 cm area around the cheek and zygomatic region. During measurement, ensure the sensor is placed flat against the skin with consistent contact pressure. The green indicator light flashes rapidly to indicate that measurement is underway; the blue indicator light indicates that measurement has been completed and data has been transmitted.
0

4. Data Recording

The software automatically records the post-measurement values of skin water content, skin oil content, temperature, humidity, and contact pressure. If a zero value, abnormal fluctuations, or probe movement during measurement occur, the corresponding result shall be discarded and the measurement re-performed. Each measurement point may be tested repeatedly, and information regarding the subject, measurement site, product used, and timestamp shall be saved simultaneously.
0

5. Efficacy Assessment

5.1 Moisturizing Effects
If, after application of the test sample, the measured values of either skin water content or transepidermal water loss (TEWL) are significantly superior to those recorded before application (P<0.05), or if the results of the test sample group are significantly superior to those of the control group (P<0.05), it is concluded that the product exhibits moisturizing efficacy at the specified testing time point; otherwise, it is concluded that the test product lacks moisturizing efficacy.

5.2 Oil Control Efficacy

If the measured skin sebum content after using the sample is significantly lower than that before use (P<0.05), or if the results of the sample group are significantly lower than those of the control group after use (P<0.05), it shall be concluded that the product exhibits oil-control efficacy at the tested time point; otherwise, it shall be concluded that the test product lacks oil-control efficacy.

frequently asked questions

1. What is the significance of measuring skin water content and skin sebum content?
Answer: An optimal balance between skin moisture and sebum levels is a hallmark of healthy skin and plays a pivotal role in protecting and maintaining skin integrity. When these two components are well-balanced, the skin exhibits a smooth, translucent texture and a natural complexion—a critical consideration from the perspective of cosmetic product development.
0
2. What are the available methods for measuring skin sebum levels?
Answer: Commonly used principles for skin sebum testing include the transmission photometer method (e.g., Sebumeter), the reflection photometer method (e.g., HydraSebum Duo), and the microbalance method (e.g., SebumScale).
3. What is the measurement principle of the HydraSebum Duo skin oil and moisture analyzer?
Answer: Sebum measurement is performed using a reflectance photometer method, in which blue-light LEDs emit light onto the skin surface; the intensity of the reflected light is then measured, and the sebum content is calculated based on the variations in reflected light intensity induced by differences in skin oil content. Skin water content is determined by measuring the electrical conductivity of the skin surface to assess the water content of the stratum corneum. Compared to other skin hydration measurement principles, this method can detect subtle variations even when skin hydration levels are relatively high.

sum up

Ultimately, the evaluation of facial oil control should not rely solely on the observation that the skin feels "refreshing after cleansing." Instead, it should assess whether changes in sebum levels can be stably and reproducibly observed, while also confirming that there is no significant decline in skin hydration. The standard T/GDCA 008-2021 serves as a framework for human testing of cleansing products regarding moisturizing and oil control capabilities, while the EEMCO guidelines provide supplementary environmental and operational requirements for hydration and electrical property measurements.

reference material

[1] T/GDCA 008-2021 "Human Testing Methods for Moisturizing and Oil-Controlting Effects of Facial Cleansing Products".
[2] Berardesca E, [2] Berardesca E, et al. EEMCO guidance for the assessment of stratum corneum hydration: electrical methods. Skin Research and Technology, 1997.
[3] Berardesca E, [3] Berardesca E, et al. The revised EEMCO guidance for the in vivo measurement of water in the skin. Skin Research and Technology, 2018.
Contact
Leave your information and we will contact you.

COMPANY

Contact us

About us

Join with us

Our Clients

Terms and Conditions

Privacy Policy

Sitemap

QUICK LINKS

Home

Products

Join with us

Blogs

FAQ

Service Request

电话